New TopShim run:

version = topshim 4.4.1
date = 2025-09-16
time = 16:06:17.570
command = topshim 

FINE SHIMMING

Parameters:

probe = Z114607_0305
solvent = DMSO
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 2.49 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 7.7 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration 2: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.8 Hz
Iteration 3: FWHM Before= 0.8 Hz, Target= 0.3 Hz, After= 0.3 Hz
Iteration 4: FWHM Before= 0.3 Hz, Target= 0.3 Hz
Shim succeeded

Lineshape quality factor (1D spectrum, spinning) = 0.437

shim changes: 
Z	-163
Z2	-24
Z3	-31
Z4	+564
Z5	-274
Z6	-2100
Z7	+506
Z8	+5100

duration = 57 sec
completed successfully
finished Tue Sep 16 16:07:14 2025


========================================================================